How to Determine Purity from Gas Chromatography

Gas chromatography (GC) is an analytical technique that separates and analyzes volatile compounds in a mixture to determine their purity.

GC is versatile and can be used to analyze a wide range of compounds, from small molecules like solvents to larger, more complex substances.

In this article, you will learn how GC works, the factors that affect separation, and how to interpret chromatograms to assess the purity of your samples.

Key Points of Article

  • Gas chromatography (GC) separates and analyzes volatile compounds to determine sample purity through differential partitioning between a carrier gas and a stationary phase.
  • GC is widely used in various fields like environmental monitoring, food analysis, forensic science, and pharmaceutical quality control for purity assessment.
  • The two main types of GC are gas-liquid chromatography (GLC) and gas-solid chromatography (GSC), each with unique applications and properties.
  • Key factors affecting GC separation include carrier gas choice and flow rate, column type and dimensions, temperature programming, and sample preparation techniques.
  • Steps to determine sample purity using GC include preparing the sample, setting up the GC instrument, injecting the sample, analyzing the chromatogram, and calculating purity based on peak areas and reference standards.
  • Birch Biotech offers high-quality LC-MS and HPLC-grade products to ensure accurate and reliable results for your lab's purity analysis needs. Contact our team to learn more.

What is Gas Chromatography?

Gas chromatography (GC) is an analytical technique used to separate and analyze volatile compounds in a mixture. It relies on the differential partitioning of compounds between a mobile phase (carrier gas) and a stationary phase (column) to achieve separation.

GC is widely used in various fields, including environmental monitoring, food analysis, forensic science, and pharmaceutical quality control. Its ability to separate and quantify individual components in complex mixtures makes it a valuable tool for purity assessment.

Example of GC analysis

One common application of GC is determining the purity of solvents like acetone. To analyze acetone purity:

  1. A small sample of acetone is injected into the GC system.
  2. The sample is vaporized and carried through the column by the carrier gas.
  3. As the sample travels through the column, its components interact differently with the stationary phase, causing them to separate.
  4. The separated compounds are detected and quantified by a detector at the end of the column.
  5. The resulting chromatogram shows peaks corresponding to each component, with the area under each peak proportional to its concentration.

By comparing the peak areas and retention times to reference standards, you can calculate the purity of the acetone sample and identify any impurities present.

Types of Gas Chromatography

Gas chromatography comes in two main types, each with its own unique properties and applications:

Gas-liquid Chromatography (GLC)

GLC is the more widely used form of gas chromatography. It employs a liquid stationary phase coated on an inert solid support inside the column. As the vaporized sample travels through the column, its components interact with the liquid phase, resulting in separation based on their relative solubility and vapor pressure.

GLC can be used to analyze a broad range of volatile compounds, from small molecules to larger, more complex substances. The choice of liquid stationary phase can be tailored to the specific analytes of interest, allowing for optimized separation and resolution.

GLC can be coupled with different types of detectors, such as flame ionization detectors (FID), thermal conductivity detectors (TCD), and mass spectrometers (MS). This flexibility enables the detection and quantification of a wide range of compounds, even at trace levels.

Gas-solid Chromatography (GSC)

Gas-solid Chromatography is a lesser-used variant of GC for purity measurements.

Unlike GLC, GSC uses a solid adsorbent material as the stationary phase. The separation of compounds in GSC is based on their relative adsorption and desorption rates on the solid surface. Common adsorbents include activated charcoal, molecular sieves, and porous polymers.

GSC is particularly useful for separating gases and low-molecular-weight compounds that are difficult to analyze using GLC. It is often employed in the analysis of permanent gases, such as oxygen, nitrogen, and carbon dioxide. Additionally, GSC is effective in separating isomers and purifying various gases.

GSC typically has lower resolution and peak capacity compared to GLC, making it less suitable for complex mixtures. Additionally, the solid adsorbents used in GSC can be sensitive to moisture and may require careful sample preparation and conditioning to maintain their performance.

Understanding the differences between GLC and GSC allows you to select the most appropriate technique for your specific purity analysis needs. While GLC is the more versatile and widely used method, GSC offers unique advantages for certain applications, particularly in the analysis of gases and low molecular weight compounds.

Benefits of Using Gas Chromatography for Purity Analysis

Gas Chromatograph has a wide range of benefits for purity analysis. Here are the primary advantages of choosing this as your preferred method of determining purity.Ā 

High Sensitivity and Resolution

Gas chromatography offers exceptional sensitivity, allowing you to detect and quantify trace impurities in your samples. The high resolution of GC columns enables the separation of closely related compounds, ensuring accurate purity assessments. Even minor impurities can be identified and measured, providing a comprehensive understanding of your sample's composition.

Versatility in Analyzing a Wide Range of Compounds

GC is a versatile technique that can handle a broad spectrum of volatile compounds. From small molecules like solvents and gases to larger, more complex substances, GC can be adapted to analyze a wide variety of samples. This versatility makes GC an indispensable tool for purity analysis across different industries and applications.

Quantitative Analysis Capabilities

GC not only separates compounds but also provides quantitative information about their concentrations. By using reference standards and calibration curves, you can accurately determine the amount of each component in your sample. This quantitative analysis capability allows you to assess the purity of your compounds with precision and confidence.

Reproducibility and Reliability

GC is known for its excellent reproducibility and reliability. When properly maintained and operated, GC systems deliver consistent results across multiple runs and users. This reproducibility is crucial for quality control and regulatory compliance, ensuring that your purity analyses are reliable and defendable.Ā 

Moreover, GC methods can be validated and standardized, further enhancing the trustworthiness of your results.

How to Determine Purity from Gas Chromatography - Supporting Image

Factors Affecting Gas Chromatography Separation

Several factors influence the separation efficiency and resolution of your GC analysis. Understanding and optimizing these parameters will help you achieve better purity determinations.

Choice of Carrier Gas and Flow Rate

The carrier gas is vital in transporting your sample through the column. Helium, nitrogen, and hydrogen are common choices, each with its own advantages. Helium provides high resolution and fast analysis times, while nitrogen is more affordable but may result in longer run times.Ā 

Hydrogen offers the best separation efficiency but requires special safety considerations. The flow rate of the carrier gas affects the speed and resolution of your analysis. Higher flow rates lead to faster elution but may compromise separation, while lower flow rates improve resolution but increase analysis time.

Column Type and Dimensions

The column is the heart of your GC system, and selecting the right one is key to successful purity analysis. Capillary columns, with their high resolution and efficiency, are the most widely used. The length, inner diameter, and film thickness of the column impact separation.Ā 

Longer columns provide better resolution but longer analysis times, while shorter columns are faster but may sacrifice some separation. Narrower inner diameters improve efficiency but reduce sample capacity, while thicker stationary phase films retain compounds longer but may cause peak broadening.Ā 

The stationary phase chemistry should be chosen based on the polarity and volatility of your analytes.

Temperature Programming

Temperature plays a significant role in GC separation. The column temperature affects the vapor pressure and elution rate of your compounds. Higher temperatures lead to faster elution but may cause peak overlapping, while lower temperatures improve separation but increase analysis time.Ā 

Temperature programming, where the column temperature is gradually increased during the run, is a powerful technique for analyzing complex mixtures. It allows you to separate compounds with a wide range of boiling points and optimize resolution.Ā 

Careful optimization of the initial temperature, ramp rate, and final temperature is necessary to achieve the best separation for your specific sample.

Sample Preparation and Injection Technique

Proper sample preparation is essential for accurate purity analysis. Impurities, matrix interferences, and non-volatile components can affect the quality of your GC results. Filtration, dilution, and extraction techniques help clean up your sample and improve injection reliability. The choice of solvent for sample dissolution is also important, as it should be compatible with your analytes and not interfere with the separation.Ā 

The injection technique, such as split, splitless, or on-column injection, depends on the concentration and volatility of your sample. Split injection is suitable for concentrated samples, while splitless injection is better for trace analysis.Ā 

On-column injection provides the best quantitative accuracy but requires a clean sample to avoid column contamination.

How to Determine Purity Using Gas Chromatography

To determine the purity of your sample using gas chromatography, follow these steps:

Step 1: Prepare the Sample and Reference Standards

Dissolve your sample in an appropriate solvent, ensuring it is compatible with your GC system and does not interfere with the separation. If necessary, filter or dilute the sample to remove impurities and adjust the concentration.Ā 

Prepare reference standards of known purity for each compound of interest, covering the expected concentration range in your sample.

Step 2: Set up the GC Instrument With Appropriate Parameters

Before you inject the sample, here are the things you should do to set up the GC instrument.

  • Select the appropriate column, carrier gas, and detector for your analysis.
  • Set the injection port and detector temperatures, as well as the column temperature program, based on the volatility and boiling points of your analytes.Ā 
  • Adjust the carrier gas flow rate to achieve optimal separation and resolution.
  • Ensure the system is properly calibrated and maintained before starting your analysis.

Step 3: Inject the Sample and Analyze the Chromatogram

Using a syringe or autosampler, inject a small volume of your sample into the GC system. As the sample travels through the column, monitor the detector signal and record the chromatogram.Ā 

Each peak in the chromatogram represents a different compound in your sample. The retention time and peak shape provide qualitative information, while the peak area or height is proportional to the compound's concentration.

Step 4: Calculate the Purity Based on Peak Areas and Reference Standards

Integrate the peaks in your chromatogram to determine their areas. Compare the retention times and peak areas of your sample to those of the reference standards.Ā 

Calculate the concentration of each compound in your sample using the calibration curves and response factors derived from your reference standards.Ā 

To determine the purity, divide the peak area of your target compound by the total area of all peaks in the chromatogram, excluding the solvent peak. Multiply this ratio by 100 to express purity as a percentage.

Gas chromatography is a powerful technique for assessing purity, as it can detect and quantify trace impurities in your sample. However, it is important to interpret the chromatogram carefully, as co-eluting peaks or overlapping impurities may affect the accuracy of your purity determination.Ā 

In such cases, using complementary techniques like mass spectrometry or tandem GC-MS can help identify and resolve these issues.

Tips for Accurate Purity Determination by GC

While GC is an ideal method of purity determination, there are certain factors that can decrease its accuracy. Hereā€™s how to avoid those pitfalls and maximize the precision of your purity determination.

Use High-Purity Solvents and Reagents

The quality of your solvents and reagents directly impacts the accuracy of your purity determination. Impurities in these materials can introduce interferences, leading to false positives or inaccurate quantitation.Ā 

To minimize these issues, always use solvents and reagents specifically designed for chromatography, such as HPLC grade or LC-MS grade products. These high-purity solutions undergo rigorous quality control testing to ensure minimal contaminants and consistent performance.Ā 

Investing in high-quality products may have a higher upfront cost but will save you time and resources in the long run by providing reliable results and reducing the need for troubleshooting.

Optimize GC Parameters for Best Separation

Achieving optimal separation is key to accurate purity determination. Take the time to fine-tune your GC parameters, such as column type, temperature program, and carrier gas flow rate, to suit your specific analytes and sample matrix.Ā 

Experiment with different stationary phases and column dimensions to find the best combination for your application. A well-optimized method will provide sharp, well-resolved peaks, minimizing the risk of co-elution and enabling precise quantitation.

Perform Regular Maintenance and Calibration

Proper maintenance and calibration of your GC instrument are essential for consistent and reliable performance. Develop a routine maintenance schedule that includes tasks such as replacing septa, liners, and columns as needed, as well as cleaning the injection port and detector.Ā 

Regular leak checks and flow rate verifications help ensure your system is operating at peak efficiency. Calibrate your instrument using certified reference standards to establish accurate retention times and response factors.Ā 

By keeping your GC system in top condition, you can minimize downtime, improve reproducibility, and have confidence in your purity results.

Validate the Method With Reference Standards

Method validation is a critical step in ensuring the accuracy and reliability of your purity determination. Use certified reference standards with known purities to validate your GC method. Analyze these standards under the same conditions as your samples and compare the results to the expected values. Calculate key performance parameters such as linearity, precision, accuracy, and limit of detection to demonstrate the validity of your method.Ā 

How to Determine Purity from Gas Chromatography - Supporting Image 2

Complementary Techniques for Purity Assessment

While gas chromatography is a powerful tool for determining purity, combining it with other analytical techniques can provide a more comprehensive understanding of your sample's composition.Ā 

These complementary methods can help you identify unknown impurities, confirm the structure of your target compound, and cross-validate your purity results.

Mass Spectrometry for Compound Identification

Coupling GC with mass spectrometry (MS) allows you to determine the mass spectrometric purity of your sample. As compounds elute from the GC column, they are ionized and fragmented in the MS. The resulting mass spectrum provides a unique fingerprint for each compound, enabling you to identify unknown peaks and confirm the identity of your target analyte. GC-MS is particularly useful for complex mixtures or when dealing with unknown impurities.

Methods like HPLC or SFC

In addition to GC, other chromatographic methods, such as high-performance liquid chromatography (HPLC) or supercritical fluid chromatography (SFC), can provide complementary information about your sample's purity. These techniques are suitable for non-volatile or thermally labile compounds that may not be amenable to GC analysis.

By comparing the purity results obtained from different chromatographic methods, you can gain a more complete picture of your sample's composition and identify any discrepancies or limitations in your GC analysis.

Spectroscopic techniques like NMR or IR

Spectroscopic techniques such as nuclear magnetic resonance (NMR) and infrared (IR) spectroscopy can provide structural information about your sample, helping you confirm the identity and purity of your target compound.Ā 

NMR spectroscopy measures the magnetic properties of atomic nuclei, yielding detailed information about the chemical environment and connectivity of atoms in your molecule. IR spectroscopy, on the other hand, probes the vibrational modes of chemical bonds, providing a characteristic spectrum that can be used for compound identification and purity assessment.Ā 

These techniques complement GC by offering orthogonal data that can help you validate your purity results and identify any impurities or degradation products.

Closing Thoughts

In conclusion, gas chromatography (GC) is a powerful and widely used technique for determining purity. It excels in analyzing volatile, thermally stable compounds with high sensitivity and can detect trace impurities at the parts-per-billion level. However, it's essential to consider whether GC is the best fit for your specific needs.Ā 

For samples containing non-volatile, thermally labile, or highly polar compounds, techniques like high-performance liquid chromatography (HPLC) or supercritical fluid chromatography (SFC) may be more suitable. Additionally, for higher accuracy and lower detection limits, advanced methods like tandem mass spectrometry (MS/MS) or high-resolution mass spectrometry (HRMS) might be necessary.

While GC provides excellent results, combining it with other analytical techniques, such as GC-MS, NMR, or IR spectroscopy, can offer a more comprehensive understanding of your sample's purity. Ultimately, the choice of analytical method should be based on your sample characteristics, purity requirements, and available resources.Ā 

Birch Biotech simplifies this process by offering the highest quality LC-MS and HPLC Grade products, ensuring accurate and reliable results for your lab. Contact our team to learn more and determine the best method for your purity analysis needs.

Sources Used for this Article:

  • sciencedirect.com/topics/agricultural-and-biological-sciences/gas-chromatography
  • scienceinfo.com/gas-liquid-chromatography-applications/
  • academic.oup.com/toxres/article/3/3/177/5575352
  • sciencedirect.com/topics/medicine-and-dentistry/thermal-conductivity-detector
  • dynamicscience.com.au/tester/solutions1/chemistry/analytical%20chem/glcms.htm
  • sciencedirect.com/topics/chemistry/gas-solid-chromatography
  • chem.libretexts.org/Bookshelves/Analytical_Chemistry/Physical_Methods_in_Chemistry_and_Nano_Science_(Barron)/03%3A_Principles_of_Gas_Chromatography/3.01%3A_Principles_of_Gas_Chromatography
  • ncbi.nlm.nih.gov/pmc/articles/PMC5206469/
  • glsciences.com/technique/technique_data/gc/basics_of_gc/p2_1.html
  • svmsl.chem.cmu.edu/caffeine/caffeine_temperature.htm
  • restek.com/global/en/articles/guide-to-gc-column-selection-and-optimizing-separations
  • chem.libretexts.org/Bookshelves/Analytical_Chemistry/Supplemental_Modules_(Analytical_Chemistry)/Instrumentation_and_Analysis/Chromatography/High_Performance_Liquid_Chromatography
  • sciencedirect.com/topics/biochemistry-genetics-and-molecular-biology/supercritical-fluid-chromatography
  • sciencedirect.com/topics/agricultural-and-biological-sciences/nuclear-magnetic-resonance-spectroscopy
  • chem.libretexts.org/Bookshelves/Physical_and_Theoretical_Chemistry_Textbook_Maps/Supplemental_Modules_(Physical_and_Theoretical_Chemistry)/Spectroscopy/Vibrational_Spectroscopy/Infrared_Spectroscopy

Disclaimer: The content provided on the Birch Biotech blog is for educational and entertainment purposes only. The information offered here is designed to provide helpful insights and advice related to laboratory practices and supplies.

Readers are advised to refer to our product-specific quality data sheets and Certificates of Analysis (COAs) available on our website for detailed information on product specifications. It is essential to handle and store all materials according to the safety guidelines and regulatory requirements applicable to your area.

While we endeavor to ensure the accuracy and relevance of the information published, it should not be used as a substitute for professional advice or official protocols. We encourage all our readers to consult their institution's guidelines, local regulations, and professional standards before implementing any practices discussed here.

Birch Biotech does not accept liability for any actions undertaken based on the information provided in this blog nor for the misuse of our products. Furthermore, Birch Biotech does not guarantee the completeness, reliability, or timeliness of the information contained on this website.

This disclaimer is subject to change at any time without notifications.

Everything for you cannabis needs banner on a colorful background
Everything for you cannabis needs banner on a colorful background